Inductively Coupled Plasma Optical Emission Spectrometer (ICP-OES)

YK-6500D

Product Category: Inductively Coupled Plasma (ICP)

The YUKE 6500D full-spectrum direct-reading ICP-OES has the excellent anti-interference, analysis and  detection  performance,  and  is  used  to  fulfill the  high-end  analysis  demands  in  research  and development as well as detection. And it can be used to carry out qualitative, semi-quantitative and precisely quantitative analysis easily, the ideal instrument for the simultaneous analysis of major, minor and trace inorganic elements.


Products Detail Technical Parameter Feature And Application Fields Of Use Details Download

Technical specifications

3.1. RF plasma source:

3.1.1: The RF power with the design of dual power supplies under full digital control has a wider power range of 500 W - 1,600 W, the adjustable power, and the stronger sample adaptability.

3.1.2:  Self-excited  RF  generator:  the  self-excited   power  supply  can  be   used  to  analyze   100% methylbenzene samples directly thanks to its high matching speed and strong power load capacity. It also  has the functions of automatic tuning and water cooling, and thus  is adaptable to the  power switching for complex sample analysis. As it has no moving part, it is resistant to either vibration or interference, which makes it more reliable;

3.1.3: iStandby mode: in this mode, the ultra-low power of 500 W for standing by is provided, and the argon consumption  is  reduced  by  over 50%; the  argon consumption during the standby  period  is smaller than or equal to 5 L/min;

3.1.4:  Observation  method of plasma: four observation  modes  are  configurable,  i. e. axial,  radial, bi-directional and simultaneous bi-directional observation modes. The adoption of vertical rectangular tubes  can  prevent  high-salinity  deposition,  extend  its  service  life,  and  reduce  the  use  of  its consumables. In axial observation, the metal cooled cone is adopted to remove tail flame, and obtain high sensibility; no configuration of any air compressor is required, saving both external fittings and consumption. In radial observation, the location of plasma observation is adjustable for the analysis 

demands of different elements, providing the stronger anti-interference capacity;

3.1.5: Intelligent attenuation: there are the functions of axial attenuation and radial attenuation, and the attenuation  of the  samples with  the  concentration within  100  times  is  realizable.  This  means  the analysis of the element that  has a high content  in the sample can be completed at the first time, requiring no repeated dilution. In this way, the preprocessing of the sample will be less difficult, and the analysis can be simplified;

3.1.6: Radio Frequency (RF): 27.12 MHz; coupling efficiency: >80%;

3.1.7: RF power stability: ≤0. 1%; RF stability: ≤0.01%;

3.2. Optical system:

3.2. 1: The thermostatic 3D optical system has a few reflections and a low optical energy loss. All optical elements are sealed in the thermal balancing optical chamber, and the host and optical chamber are isolated thermally, which can resist the external ambient temperature changes well. Optical chamber: precise; kept at the constant temperature of 36±0. 1; argon removing;

3.2.2: Stable, efficient fully-fixed echelle optical-splitting system: with no moving parts but high vibration resistance, stability and reliability;

3.2.3: Wavelength range: 165nm - 900 nm, with full wavelengths covered;

3.2.4: Monochromator: focal length=380 mm; with the total reflection imaging optical path and quartz prism 2D dispersive system;

3.2.5: Echelle grating: 87 L/mm, 63.5° blazing; the larger the number of ruled grating lines is, the higher the resolution will be;

3.2.6: Wavelength correction: For each time of ignition, only spectral lines of C, N and Ar are used for  the automatic correction of the spectral position to ensure the correctness of analysis wavelengths, and no wavelength correction solution is required;

3.2.7:   Patented   full-spectrum   real-time   calibration   technique   (FSC):   the   interference-free   Ne characteristic spectral line is utilized to fulfill the real-time calibration of the fine spectrum offset, realize the optimal spectral integral, ensure long-term stability, and eliminate the effect of the spectral drift on measurement well.

3.2.8: Purging optical chamber: for the determination of the wavelengths shorter than 189 nm, argon is selected to purge the optical path, and no vacuum pump is required to prevent the vacuum return oil   that will contaminate the optical chamber;

3.2.9: Stray light: ≤2.0 mg/L (10,000 mg/L Ca solution is determined at the location of As 188.980 nm);

3.2.10: Optical resolution (FWHM): ≤7 pm @ 200 nm;

3.3. Detector: special large-area array ECCD detector with pixel-level refrigeration

3.3.1:  Detection  unit:  large-area  array  CCD  detector  with  the  1,024* 1,024  pixel  size,  in  a  single exposure;

3.3.2: Imaging size: 25.4 mm×25.4 mm large-area array imaging and photosensitive units; the large picture element of 24 um × 24 um is adopted to fulfill high-sensitivity response; as the overall breadth is large, both the high resolution and a wider spectral range can be obtained;

3.3.3: Pixel-level refrigeration: the TEC refrigeration part that is encapsulated in the sensor acts on the pixel  element  directly, and the  cryogenic temperature  is  higher  than  -10 , which  can  effectively prevent the CCD surface condensation while require no gas purging protection;

3.3.4: Anti-saturation overflow: the back relief anti-overflow protection is designed for each pixel to eradicate thoroughly the problem of spectrum saturation overflow, and eliminate the concern about the effect of spectrum saturation on the adjacent spe

Sample introduction system & Software performance

3.4. Sample introduction system:

3.4.1: Sampling system: the quartz concentric nebulizer, quartz swirling flow atomizing chamber and detachable rectangular tubes are the standard accessories;

3.4.2: Argon is optional for on-line dilution. The high-precision MFC controls argon to dilute efficiently the high-salinity samples with the saltness over 10%, and samples are introduced directly.

3.4.4: The sampling system resistant to high salinity or HF, the organic sampling system and hydride generators are optional;

3.4.5:  Peristaltic  pump:  12-roller 4-channel  peristaltic  pumps are  adopted, and the  pump speed  is adjustable from 0 rpm to 125 rpm automatically and continuously. When stable sampling is ensured, the  simultaneous operation  of  sampling  tubes,  internal  standard  tubes,  waste   liquid  tubes  and 

accessory reagent tubes (the hydride generator) can be supported, which is conducive to the analysis of complex samples;

3.4.6: Control of the gas channel: the precise mass-flow controller is adopted to control the flow of multiple-channel gases, and at  most 5 channels can  be controlled,  including the atomization gas, auxiliary gas, cooling gas and additional expandable gases (O2  and Ar), with the accuracy of 0.01 L/min.

3.5. Software performance:

3.5.1: Thanks to the graphical operation  interface, the software operation  is visual and easy. The interface provides the functions of qualitative, semi-quantitative and quantitative analyses;

3.5.2: The method library management software based on classification and versions is conducive to the management, maintenance and preservation of methods; some standard methods that are built in the software can facilitate analysis efficiency;

3.5.3: It has the function of full-spectrum data acquisition, and can record the spectrum lines of all elements. The data can be stored safely, and the functions of both analysis data storage and retrieval are supported for reanalysis in the future;

3.5.4: There is a spectral line library with over 50,000 spectral lines, and at least 30 pixels for each spectral line can be selected for measurement;

3.5.5: It has the function of full-spectrum acquisition, and a complete full spectrum can be obtained from the software to know about the sample spectrum and spectrum interference status;

3.5.6:  It  has  multiple  interference-correction  methods  and  the  function  of  real-time   background subtraction, such as the standard comparison method, internal standard method, interfering element correction coefficient method (IEC) and standard addition/curve methods, which enrich users’ multiple means of analysis and study;

3.5.7: It has the function of instrument calibration, and supports the functions including rectangular tube collimation and light source optimization, which are conducive to users’ routine maintenance. It also provides the visualized monitoring of instrument operation status;

3.5.8:  It  provides  the  login  password  protection,  multi-level  operation  permission  setting,  network security management, and permanent storage of historic records;

3.5.9: It has the visualized observation module of rectangular tube flames;

3.5.10: There are both English and Chinese versions of the software;

3.5.11: It has the function of network-based remote service, and has a built-in remote service assistant 

for remote diagnosis; the 4G  Network  Data Connection Technology Service  Department fulfills the remote diagnosis and maintenance of instruments;

3.5. 12: The software design is in complete compliance with the 21 CFR Part 11- regulation of electronic signature; the software provides the three-level administrative authority and the function of audit trailing, conforming to the regulation requirements, such as 3Q certification;

3.5.13:  It  can  integrate the software operations of the  automatic analysis  instrument  platform and on-line analysis instrument platform.


Analysis performance & Instrument interface

3.6. Analysis performance:

3.6. 1: Analysis speed: about 200 spectral lines per minute;

3.6.2: Sample consumption: only 2 ml;

3.6.3: The determined linear dynamic range of spectral lines: ≥105  (determined by using Mn 257.6 nm; the relevant coefficient ≥0.999);

3.6.4: Precision: determine 1 ppm or 10 ppm multi-element mixed standard solutions; the RSD from the repeated 10 times of determination: ≤0.5%;

3.6.5: Stability: determine 1 ppm or 10 ppm multi-element mixed standard solution; the RSD, for the long-term stability over an 8-hour period: ≤1%;

3.6.6: Detection limit: (unit: ug/L, the element conforms to the specifications of JJG768-2005)

 

Zn213.856

Ni231.604

Mn257.610

Cr267.716

Cu324.754

Ba455.403

0.5

1

0.5

1

1

0.1

3.6.7: It can implement qualitative, semi-quantitative and quantitative analysis of any spectral line for the analysis element; it also supports the methods including the internal standard method, standard addition method and interfering element correction;

3.6.8: Preheating time: the duration from the standby period to the time when the plasma is ignited is shorter than 5 minutes.

3.7. Instrument interface (including coupling analysis software):

3.7.1:  It  can  fulfill  the  coupling  of  the  automatic  digestion  sample  processing  workstation  (in  the digestion  system, the  graphite-digestion  electro-thermal  mode  is  adopted  to  realize  the  automatic digestion, isochoric and sampling operations, and the simultaneous processing of the samples at over 48 sites is supported)

3.7.2 It is equipped with the analysis system interface for direct organic sample introduction (it provides the function of sample refrigeration, with the temperature below -15);

3.7.3  It is equipped with the on-line digestion module interface (supporting the functions of on-line automatic  electric  heat  digestion  and  concentration  control  as  well  as  the  automatic  cleaning  of sampling pipelines);

3.7.4 It is equipped with the on-line analysis system interface for hydride generation (supporting the functions of the automatic addition and concentration control of reactive liquid as well as the automatic cleaning of sampling pipelines);

3.7.5  It can be connected to the  ion exchange workstation for valence analysis or the analysis of low-concentration enrichment high-matrix removal;

3.7.6 It can be connected to the automatic sample injector for analysis (supporting the configurability of over 240 sample sites);


After-sales service

Project building is a complex, long-term technical service. In order to fulfill the system service earnestly, the following principles are prepared as the main principles of instruction:

 Principle of response timeliness

The  Technical  Service  Department  and  Customer  Service  Center  of  the  YUKE  Technology  will undertake the quick response of the construction of this project, and provide technical consultation service in time.

 Principle of normative service

Our implementation engineers, technical supporting engineers and maintenance engineers do not only have professional technical skills, but also have been certified by the training about the company’s customer  service  specifications.  They  provide  customer  service  in  strict  accordance  with  our company’s customer service specifications.

 Principle of high efficiency in solving problems

The principle of the high efficiency in solving problems refers to the quick positioning and problem solving through the professional skills of the site supporting engineers.

 Process of customer service response

“Customer-centricity”  is the  important  business  philosophy of the  YUKE Technology.  In order to practically implement this philosophy, and to ensure the timely response to and standard service for customers’ problems, the company has established a scientific, reasonable customer service response system. The Customer Service Center of the YUKE Technology will provide technical consultation services for customers.

 Periodic preventive system maintenance

The status report is required for periodic inspection.  In the report, there shall be the detailed error recording of each part of hardware, and measures can be adopted as soon as possible accordingly to eliminate  hidden  faults.  The  engineers  of  our  companys  Technical  Center  periodically  implement remote login by the Internet to fulfill 24-hour automatic reporting and tracking against problems. Then, these engineers can provide timely maintenance and repair based on the relevant results. Periodic equipment testing and inspection are implemented to find out hidden faults, and to eliminate them as soon as possible.

 Rapid site service

If there is any system failure diagnosed in the customer’s system that cannot be solved by means of telephone/email,  remote  dial-in  analysis,  etc.,  our  site  engineers  will  immediately  take  along  the corresponding system tools and software, and head for the site for emergent maintenance. We promise that we will  respond to  customers within  24  hours  after the system failure,  and will  arrive  at  the customer’s site within 3 working days in case of any site maintenance.

 Periodic visit and communication

During  the  operation  of  the  customers  system,  we  will  adopt  the  method  of  designating  special personnel to  implement visits and communication  regularly or  irregularly to  investigate the  project implementation and maintenance, listen to customers’ opinions, solve customers’ problems at site, as well as test and optimize systems at site. In this way, we can find out system problems or hidden faults, eliminate  hidden  faults  in  advance,  ensure  safe,  stable  system  operation,  and  fulfill  the  quality assessment of the system.

The above customer service response methods are not isolated or unrelated to each other. However, supported by the organizational structure of the Customer Service Center of the YUKE Technology, multiple flexible service modes permeate into, and are combined closely with, each other, forming a complete, unified fault response system for customers’ service. This system has been experienced practically in many large-scale working projects, and has been considered as the efficient, feasible and optimal fault response method.



Work environment

Voltage: 220V AC ± 10%

Room temperature: ambient temperature (10-30)

Relative humidity: (20-80)%RH